Turbidity Meter

Turbidity is measured according to EN ISO 7027 by nephelometric means (90° scattered light). The infra-red light-source permits measurement of coloured and colour-free samples. The automatic measurement range detection facility (Autorange) enables di

Turbidity Meter

Contacts and Location

Description

Turbidity is measured according to EN ISO 7027 by nephelometric means (90° scattered light). The infra-red light-source permits measurement of coloured and colour-free samples. The automatic measurement range detection facility (Autorange) enables direct turbidity measurement from 0.01 to 1100 NTU with an accuracy of ± 2% up to 500 NTU and ± 3% thereafter.

Specification

Display: Graphic Display (7-line, 21-characters). Light source LED λ=860 nm and photo sensor amplifier in protected cell compartment Principle: Nephelometric (Non Ratio). Range: 0.01 – 1100 NTU. Resolution: 0.01 – 9.99 NTU = 0.01 NTU, 10.0 – 99.9 NTU = 0.1 NTU, 100 – 1100 NTU = 1 NTU. Accuracy: ± 2 % of reading or ± 0.01 NTU from 0.01 to 500 NTU, ± 5 % of reading from 500 to 1100 NTU. Repeatability: ± 1 % of reading or ± 0.01 NTU which ever is greater Operation: Acid and solvent resistant touch-sensitive keyboard with integral beeper as acoustic indicator. Auto off: 20 minutes after last function, 30 seconds acoustical signal before switching off. Charging time: approx. 10 hours. Dimensions: approx.265 x 195 x 70 mm (unit), approx. 440 x 370 x 105 mm (case) Weight (unit): approx. 1000 g. Storage capacity: ca.1000 data sets.