High performance liquid chromatograph, SHIMADZU

Dual-plunger parallel-flow solvent delivery module. It offers improved accuracy and sensitivity in high-performance liquid chromatography.

High performance liquid chromatograph, SHIMADZU

Contacts and Location

Contact 1 Anita Kalniņa
Contact 2 Pēteris Mekšs
Enquire about this equipment
Organisational Unit University of Latvia
    Faculty of Chemistry
        Department of Physical Chemistry - Latvia, Rīga

Description

Dual-plunger parallel-flow solvent delivery module. It offers improved accuracy and sensitivity in high-performance liquid chromatography.

Specification

  • Constant flow and constant pressure can be obtained
  • Flow rate range:
    • 0,0001-5 mL/min (1,0-40 MPa)
    • 5,0001-10mL/min (1,0-20 MPa)
    • ± 1% or ± 0,5 μL/min [whichever is the greater] (0,01-2 mL/min)
    • ± 2% (2-5 mL/min)
  • Pressure range: 1,0-40 MPa [0,1 MPa steps]
  • Pressure accuracy: ± 10% or 1,0 MPa, whichever is the greater
  • Suction filter: 10 μm
  • Up to 320 steps can be time programmed with flow rate, pressure, EVENT functions, LOOP
  • Operating temperature range 4-35oC
  • 2 or 3 solvents can be mixed
  • Gradient profile: step, linear and exponential functions possible at multiple levels
  • Max. 400 steps per program can be used with high-pressure gradient system controlled by system controller
  • program duration: 0,01 – 999,99 minutes (in 0,01 minutes steps)