Helios Spectrophotometers’ superior optical system provides stability, precision and linearity. All units feature angled casing, sealed keypad to protect from spills, optional printer, plug-and-play self-aligning lamps, large LCD, VGA graphics, full-
Specification
Wavelenght Calibration Method and Values: Filter Value 279.3; 360.8; 453.6; 573.0; 807.1 nm. Measured Value 278.2; 360.8; 453.0; 574.0; 807.0 nm. Absorbance accuracy Methods and Values: Filter Value 0.1030; 0.4493; 0.8884; 1.9504; 2.4356; 2.8726. Measured Value 0.1022; 0.4468; 0.8861; 1.9495; 2.4313; 2.8694
Services
Manufacturer
Thermo Spectronic (manufacturer's website)
Model
Heλios Gamma, Delta
Portability
Portable
Access Level
Full Access
Training
Training is required to use this equipment and we can arrange this if needed.
Calibrated
Yes, this equipment is calibrated.
Last Calibration: 16th August, 2006