Accred­ited EMC test­ing, EMC pre-​compliance test­ing and EMC trou­bleshoot­ing to EMC com­pli­ance test­ing.

Contacts and Location

Description

Accred­ited EMC test­ing, EMC pre-​compliance test­ing and EMC trou­bleshoot­ing to EMC com­pli­ance test­ing.

Specification

The equipment was set up as per the test configuration to simulate typical usage per user’s manual. When the EUT is a table top equipment, a wooden turntable with a height of 0,8m is used which is placed on the ground plane. When EUT is floor standing equipment, it is placed on the 0,1m insulation support. Auxiliary equipment and/or support equipment, if needed was placed as per EN 55022 recommendations. All input/output cables were positioned to simulate typical usage as per EN 55022. The EUT was connected to AC mains 230V/50Hz under the turntable shucko type socket, all other equipment was connected to the other shucko type socket under the turntable. The antenna was placed at 3m away from EUT. Antenna height was changed in range 1-4m and EUT rotation angle in range of -1800 to 1800 maximize measured emissions.