Electrostatic discharge immunity

Accred­ited EMC test­ing, EMC pre-​compliance test­ing and EMC trou­bleshoot­ing to EMC com­pli­ance test­ing.

Contacts and Location

Description

Accred­ited EMC test­ing, EMC pre-​compliance test­ing and EMC trou­bleshoot­ing to EMC com­pli­ance test­ing.

Specification

The equipment was set up as per the test configuration to simulate typical usage per user’s manual. When the EUT is table top equipment, it is placed on table 0,1m above ground reference plane. When EUT is floor standing equipment, it is placed on the 0,1m insulation above the ground reference plane. Electrostatic discharges are applied as contact discharge and air discharge, discharge to vertical and horizontal coupling plane. The discharges are applied only to such points and surfaces of the EUT which are accessible to personnel during normal usage. Test is performed as single discharges on preselected points at least ten single discharges on both polarities. Between successive discharges a time interval of 1s is used. In case of contact discharge the tip of discharge electrode touch the EUT before the discharge switch is operated. In case of air discharge, the round tip of the discharge electrode is approached as fast as possible (without causing mechanical damage) to touch the EUT discharge switch is operated before the tip is approached. At least ten single contact discharges are applied to horizontal and vertical coupling plate.

Service fields

analyse -
measurement -
testing -