Contacts and Location
Enquire about this equipment | |
Organisational Unit |
Riga Technical University Faculty of Natural Sciences and Technology Institute of Biomaterials and Bioengineering - Latvia, Rīga |
Description
Sample preparation involves the removal of surface contaminants by heating the sample under vacuum.
Services
To more effectively prepare multiple samples and to fully utilize the high throughput capabilities of the Autosorb-6B, and Quadrasorb, Quantachrome has created the Autosorb degasser.
The Autosorb DEGASSER has six preparation stations side-by-side for easy access. Each station has its own heating controller for independent operation. Each controller has a novel 'ramp' capability that gently increases heat at a selected rate. Individual digital heat settings and temperature metering assure accurate and repeatable degassing conditions. The heating of each sample is terminated manually or by a timer.
The rate of the initial evacuation can be limited to prevent elutriation (removal) of fine particles. After degassing, each sample cell is filled to ambient pressure with any dry, non-adsorbing to protect the sample from air contamination during transfer to your gas sorption analyzer.
Degas method: vacuum. Sample stations: 6. Max temp.: 350oC. Cold trap: yes. Timer: yes. Ramp controller: yes. Evacuation rate: adjustable. Dimensions: (H)74cm x (W)77cm x (D)40cm. Weight: 72 kg.