Contacts and Location
Enquire about this equipment | |
Organisational Unit |
Riga Technical University Faculty of Electrical and Environmental Engineering Institute of Energy Systems and Environment - Latvia, Rīga |
Building | Riga, Azenes,12 k-1 |
Room | 615 |
Description
The Thermal Microclimate HD32.1 instrument is manufactured by Delta Ohm Srl and it allows studying, measuring and controlling the Microclimate in the workplace, in compliance with the following standards: ISO 7726; ISO 7730; ISO 7243; ISO 7933; ISO 11079; ISO 8996
Specification
Dimensions (Length x Width x Height): 220x180x50 mm
Weight: 1100 g (with batteries)
Operative temperature: -5 … 50°C
Storage temperature: -25 … 65°C
Operative Relative Humidity: 0 … 90% RH non condensing
Mains adapter (code SWD10): 12Vdc/1A
Batteries: Four 1.5V batteries size C-BABY
Autonomy With temperature and RH probes: 200 hours with 7800mAh alkaline batteries
Accuracy: 0.5 hPa
Resolution: 0.1 hPa
Response: time 1 s
Pt100 measuring range: -200…+650°C
Resolution: 0.01°C in the range 199.99°C 0.1°C outside this range
Accuracy: 0.01°C in the range 199.99°C 0.1°C outside this range
Temperature drift @20°C: 0.003%/°C
Drift after 1 year: 0.1°C/year
Measuring range: 0…100%RH
Resolution: 0.1%RH
Accuracy: 0.1%RH
Temperature drift @20°C: 0.02%RH/°C
Drift after 1 year 0.1%RH/year
Input for probes with SICRAM modul:e 8 male 8-pole DIN 45326 connectors
67600 storages of 8 inputs each
Services
The instrument can perform the following measurements: Globe temperature; Natural wet bulb temperature; Ambient temperature; Relative Humidity; Air velocity; Air temperature at the height of the head (1,7m subject standing; 1,1m subject sitting); Air temperature at the height of the head (1,7m subject standing; 1,1m subject sitting) or abdomen (1,1 m subject standing; 0,6m subject sitting) or ankles (0,1 m) or at the floor level.
Funding Source
Other | - |
---|
Manufacturer
Delta Ohm Srl
Model